2 Menu Item Selection
Reflex HC Series
35
Changing the Calibration Mass to be Used
Pressing the
key allows changes to the weight value.
Modify the value using the
key and
key, then press
the
key. To interrupt modification, press the
key.
(4)
Load the indicated calibration weight and press
the
key.
(5)
Shortly, zero display flashes. Unload the weight from
the pan and press the
key.
is displayed briefly to indicate completion of span
calibration.
2.5.3
Calibration Check Using External Weights
(1)
Verify that the balance is in mass display and unload the
sample from the pan.
(2) Press
the key once to display
.
(3) (If
is not displayed, select Menu Map item
[4].)
(4) Press
the
key.
(5)
The value of the correct calibration mass to be loaded is
displayed and flashes.
Changing the Calibration Mass to be Used
Pressing the
key allows changes to the weight value. Modify
the value using the
key and
key, then press the
key. To interrupt modification, press the
key.