85
Command
Name
Value
LF?
List File Name
Active selected memory
LFN?
List Memory (File) Number
Active selected memory
LF <memory
number
>?
List File Name by memory
number
memory number
= 1-50
LP?
List Prompt
Active selected Step
LP <
step number
>?
List Prompt by step number
step number
= 1-30
LS?
List Step Parameters
LS <
step number
>?
List Step Parameters by step
number
step number
= 1-30
TD?
Read the active data being displayed on the LCD display while the test is in process. Will also
read the last data taken when the test sequence has completed. Each parameter is separated by
commas and includes step number, test type, test status, and metering. The syntax for this
command response is {step, test type, status, meter 1, meter 2, meter 3}. ACW test displays 4
meters. Each meter will contain only the value and not the units. In the case of DCW current
where both uA and mA are used on the display, the command response will always indicate the
current in uA for example 2.0mA will respond with 2000 for 2000uA.
RD <step number>?
Read the results for an individual step. The step number is the actual step number that has been
saved within the file, not the order of which the steps were executed. For example if the test
was executed starting from step 3 and ending with step 5 then the first step test results will be
found in location 3 not in location 1. Each parameter is separated by commas and includes step
number, test type, test status, and metering. The syntax for this command response is {step,
test type, status, meter 1,meter 2,meter 3}. ACW test displays 4 meters. Each meter will
contain only the value and not the units. In the case of DCW current where both uA and mA
are used on the display the command response will always indicate the current in uA for
example 2.0mA will respond with 2000 for 2000uA.
RR?
Read the remote Reset input signal. When the remote reset has be activated by closing the
contacts the query will return a value of 1 to indicate the instrument is being Reset.
RI?
Read the remote Interlock input signal. When the remote Interlock has be activated by opening
the contacts the query will return a value of 0 to indicate the instrument is in the Interlock state
and will not be able to generate output voltage or current.