
24
CONTINUITY TEST MODE
Output Current
DC 0.1A
±
0.01A
Total Resistance*: 0.00-33.0
Ω
DC 0.01A
±
0.001A
Total Resistance*: 31.0-330
Ω
DC 0.001A
±
0.0001A
Total Resistance*: 310-2000
Ω
Resistance Display
Range 1:
0.00 – 19.99
Ω
Resolution: 0.01
Ω
Accuracy:
±
(1 % of r 0.05
Ω
)
Range 2:
20.0 – 199.9
Ω
Resolution: 0.1
Ω
Accuracy:
±
(1 % of r 0.2
Ω
)
Range 3:
200 – 2000
Ω
Resolution:
1
Ω
Accuracy:
±
(1 % of r 2
Ω
)
Maximum and
Minimum Limits
Range 1:
0.00 – 99.99
Ω
Resolution: 0.01
Ω
Accuracy:
±
(1% of 0.05
Ω
)
Range 2:
100.0 – 999.9
Ω
Resolution: 0.1
Ω
Accuracy:
±
(1% of 0.2
Ω
)
Range 3:
1000 – 2000
Ω
Resolution:
1
Ω
Accuracy:
±
(1% of 2
Ω
)
(Max Limit: 0 = OFF)
Dwell Timer
Range:
0.0, 0.3 – 999.9 sec (0 = Continuous)
Resolution: 0.1
sec
Accuracy:
±
(0.1% + 0.05 sec)
Milliohm Offset
Range:
0.00 – 10.00
Ω
Resolution: 0.01
Ω
Accuracy:
±
(1 % of r 0.05
Ω
)
* - Total Resistance includes resistance of Test leads, Fixture, and DUT.
Summary of Contents for HYPOTULTRA III 7620
Page 7: ......
Page 8: ......
Page 130: ...122 configure it through the drive s properties in Windows ...