28
3.2.2. Rear Panel Controls
1. BUS INTERFACE:
Standard connector for interconnection to the RS-232 Bus interface.
Optional IEEE 488 interface may be substituted for the RS-232.
2. CALIBRATION BUTTON:
To put the instrument into the calibration mode, push this
button and turn on the power switch simultaneously.
3. REMOTE SIGNAL INPUT:
9-Pin D subminiature male connector for remote control of
TEST, RESET, and INTERLOCK functions, as well as remote memory selection.
4. REMOTE SIGNAL OUTPUT:
9-Pin D sub-miniature female connector for monitoring
PASS, FAIL, RESET, and PROCESSING output relay signals.
5. SCANNER CONNECTOR:
For connection of optional Scanning External Matrix.
6. HIGH VOLTAGE OUTPUT JACK:
Connector used to attach the High Voltage test
lead, adapter box high voltage lead, or test fixture high voltage lead to the instrument. The
connector is recessed for safety.
7. RETURN OUTPUT JACK:
Connector used to attach the Return test lead or test fixture
return lead to the instrument. This connection provides the return current path for the high
voltage and continuity tests.
8. CONTINUITY OUTPUT JACK:
Connection for the low current return lead or three-
prong receptacle adapter box used for continuity testing.
9. FUSE RECEPTACLE:
To change the fuse, unplug the power (mains) cord and turn the
fuse receptacle counter-clockwise. The fuse compartment will be exposed. Please replace
the fuse with one of the proper rating.
10. SCANNER OUTPUTS:
Optional scanner matrix that provides 8 HV/Continuity/Return
connections. Please refer to the Options section of this manual for additional connection
information.
1
10
2
4
3
6
5
12
11
9
8
7