Ramp DN
: A length of time that is allowed for the test voltage to decay from set point to 0.
Arc Sense
: During Hipot testing some low current arcing may be allowable. Arc sense is a
maximum allowable threshold for arcing.
The numbers corresponding to the different arc sensitivity levels are 0 through 9, 1 meaning the
maximum threshold of allowable arcing, 9 meaning the minimum threshold of allowable arcing
and 0=OFF.
Frequency
: This parameter is available in AC tests only and is selectable between 50 and 60Hz.
Continuity
: This function checks for a connection between the cont. check and return lead.
Connect
: This function when turned ON will connect or link the current step to the next step.
When step 30 connect is turned “ON”, it will be linked to the first step of the next memory. For
example, memory 1, step 30 will be connected to memory 2, step1. It is possible to connect all
1500 steps together when the connect parameter of all steps has been turned “ON”.
Scanner or EX Scanner
: (This parameter will only be seen on units equipped with a scanner).
This multiple parameter allows for set up of the scanner channels. The three different selectable
scanner states are L (scanner channel set to the return point), H (scanner set to the high voltage
point) and O (OFF).