62
5.4. Performing a Test
1. As instructed in section 4.2., select a memory and step that is suitable for the test you would
like to perform.
2. Attach the appropriate load or DUT to the instrument (refer to section 5.1 for instrument
connections).
3. Press the TEST button.
4. The instrument will now perform the test or connected sequence of tests. If the Test is
started from any other step than 01, when you push Reset or Test buttons, it will always
return to the originally selected step.
5. If a prompt is embedded within a step, the test will pause at that step and display the prompt
as a pop-up message. An example of a prompt will appear as follows:
A C W T o t a l S e t t i n g s M e m N a m e * *
M e m o r y
S t e p
E d i t
< m o r e >
6. In order to clear the prompt and continue the test, push the Test button.
7. During a test, the appropriate metering will be display for the type of test that is being
performed (Refer to section 5.3.). At the end of the test sequence, the top field of the display
will indicate PASS if all the tests have completed successfully or indicate FAIL if any of the
tests did not complete successfully.
5.4.1. Manual Voltage Adjustment
The test voltage may be adjusted up and down during a test by pressing the “+ and –“ soft keys.
A C W T o t a l D w e l l
M 5 0 - 8 _
P
2 . 3 s 1 0 . 0 0 m A R
+
_
When the Lock parameter is selected “On” in the system menu, manual voltage adjustment is
disabled. Therefore, you will not see the “+” and “-” symbols next to the bottom soft keys.
C O N N E C T T H E T E S T L E A D S
T O T H E D U T
P R E S S T E S T T O C O N T I N U E
Press the + and - soft key to adjust
the test voltage.