6.2.1. Continuity Verification, Models 7620 and 7650
From the Verification screen, press the Up or Down arrow keys to select “Continuity”, then
press the ”Select” soft key. An instructional prompt will now be displayed. The prompt will
appear as follows:
C O N T I N U I T Y V E R I F I C A T I O N
O p e n t h e D U T c o n n e c t i o n b e t w e e n C o n t .
C h e c k a n d R e t u r n . U n p l u g H V f r o m t h e
I n s t r u m e n t .
P r e s s T E S T t o b e g i n V e r i f i c a t i o n .
P r e s s R E S E T t o c a n c e l .
Follow the instructions given in the prompt and then press TEST to begin the verification
process.
At the end of the Verification process, a message will appear indicating the outcome of the
process. If the instrument passes the Verification, (test failure, indicating the fail detectors are
working properly) the RESET button will illuminate, the alarm will sound, and the following
message will appear:
C O N T I N U I T Y V E R I F I C A T I O N
V E R I F I C A T I O N O K
C o n t i n u i t y f a i l u r e d e t e c t e d .
P r e s s R E S E T t o c l e a r m e s s a g e a n d r e t u r n
t o t h e V e r i f i c a t i o n m e n u .
If the instrument fails the Verification (test pass, indicating the fail detectors are not working
properly) the following message will appear:
C O N T I N U I T Y V E R I F I C A T I O N
V E R I F I C A T I O N E R R O R
C o n t i n u i t y f a i l u r e n o t d e t e c t e d .
C h e c k l e a d s o r c a l l 1 - 8 0 0 - 8 5 8 - 8 3 7 8
T o r e t e s t C o n t i n u i t y , p r e s s T E S T .
P r e s s R E S E T t o c a n c e l .