scanned. Do not touch the DUT at any time when using this feature in
order to avoid potential shock or serious injury.
To completely enable this feature, the user must name the desired test file for a particular
product the exact alpha-numeric code that is on the product’
s barcode label. For example,
if Product A has barcode “123456789”, then the test file that the user would like to run
when testing Product A should be named “123456789”. Upon scanning the barcode, the
HypotULTRA III will immediately execute the test associated with that barcode. The test
file name is limited to 10 characters. However, if the user names a test file with the
maximum 10 characters, this function will still initiate a test when a product’s barcode
begins with those first 10 characters even if the barcode has more than 10 characters.
When using certain features of barcoding, the HypotULTRA III’
s response to the TD?
and RD x? commands will differ slightly to an HypotULTRA III with the standard RS232
interface. For all types of tests (ACW, DCW, IR, CONT) two fields are added to the end
of the standard response when the Barcode I/P setting is set to SERIAL#, PRODUCT# or
SER/PROD. The first field contains the Serial Number information and the second field
includes the Product Number information. Both fields are included regardless of which of
these three modes are selected. The Data Storage Card will simply substitute a “0” for the
field if it is not applicable to the setting. For example, if a user had their Barcode I/P
setting set to SERIAL#, and scanned a Serial Number with the value “123456789”, the
TD? response for an ACW test could be:
01,ACW,Pass,1.24,1.000,0.900,1.0,123456789,0
Note that there is a “0” in the Product Number field because the Barcode I/P setting is
SERIAL#.
When the Barcode I/P setting is RUN FILE or OFF, these fields are not included in the
TD? and RD x? responses.
Use the “Change” soft key to select the Barcode I/P. Press the ENTER key to accept the
new setting or the EXIT key to cancel and return to the original setting.