80
Command
Description
Value
SAC
Step Add CONT test
ADD
<
test
,
p1,p2,p3… >
Add all parameters of a
test
SP <
prompt message
> Step Prompt Create
prompt message
= Valid ASCII (1)
maximum 32 characters
SP
Step Prompt Delete
SF {1|0}
Step Fail Stop
1=On, 0=Off
(1) "Valid ASCII" is the character set that is available from the front panel LCD user interface.
Consisting of upper case alphabet (A-Z), numbers (0-9) and decimal point (.), asterisk (*), dash
(-), under bar (_), tilde (~) and space (SP).
FL <memory number>
Load a file by memory number from non-volatile memory into random access memory RAM.
FN < file name>
Creates a new file name for the active memory loaded into RAM.
SS <step number>
Selects the active selected step to load into RAM. The step must first be selected before any
specific parameters can be edited.
SAA, SAD, SAI, SAC
These commands add the appropriate test type within the memory at the step location that has
been selected. The parameters of the previous test type will be deleted and the default values
for the new test type will be recalled. If the same test type is selected that already exists, the
default values will replace the previous parameters.
ADD <test,p1,p2,p3… >
This command edits all parameters in a step. Parameters will be edited at the step location that
has been selected. See the command summary tables to see the specific test type for each of
these commands
The parameter <
test>
indicated the test type. The values ACW, DCW, IR, or CONT must be
used. The parameters <
p1,p2>
etc. indicate the individual settings for each test. All parameters
must be included with the command and should appear in the same order that is shown in the
table below. Also, like the individual parameter editing commands, the unit should not be
included with the value. Only the numeric value should be included in the command string.
When the scanners are being used they should be appended to the end of the string, with the
internal scanner first if installed, followed by the external scanner if connected the rear panel
scanner control port. The list of parameters can also be found in the default parameters section
of the manual, or refer to the Test Parameter Editing commands section for the proper values.