79
All commands that end with a question mark (?) are query commands and required an IEEE-
488 read command to retrieve the data from the device's output buffer.
8.9.1. Test Execution Commands
The following commands are used to control actual output voltage and current from the
instrument. Please observe all safety precautions.
Command
Description
TEST
Execute a Test
RESET
Abort a test in Process or Reset Failures
SAO
Set Auto-Offset
SACG
Set Auto-Charge-LO
TEST
Starts the test sequence at the selected step loaded into memory (RAM).
RESET
Stop or abort a test. Also used to reset a latched failure condition.
SAO
Set the offset for the Continuity test. The cables and any test fixture should be connected
before executing the command. This command will perform an actual test and all safety
precautions should be observed when using this command.
SACG
Set the Charge-LO parameter for the DCW or IR test. The cables and any test fixture should be
connected before executing the command. The test parameters that are set for the step will be
used when performing the auto setting. This command will perform an actual test and all safety
precautions should be observed when using this command.
8.9.2. File Editing Commands
The following commands are used to create or modify Test Setup Files.
Command
Description
Value
FL <
memory number
> File
Load
memory number
= 1-50
FN <
file name
>
File Name
file name =
Valid ASCII (1)
maximum 11 characters
SS <
step number
>
Step Select step
number
step number
= 1-30
SAA
Step Add ACW test
SAD
Step Add DCW test
SAI
Step Add IR test