21
DIELECTRIC WITHSTAND TEST MODE
DC
Range 1:
0.0 – 999.9
µ
A for 7650 only
Resolution: 0.1
µ
A
Range 2:
1000 – 10000
µ
A for 7650 only
Resolution: 1
µ
A
Accuracy:
±
(2
%
of setting
+
2 counts)
Arc Detection
Range:
1 – 9
Failure Detector
Audible and Visual
Voltage Display
Range:
0.00 – 5.00 KV Full Scale
Resolution: 10
volts
Accuracy:
±
(2
%
of setting
+
2 counts)
Current Display
Auto Range
AC Total Range 1:
0.000mA – 3.500mA
Resolution: 0.001mA
Range 2:
3.00 – 30.00 mA
Resolution: 0.01
mA
Accuracy:
±
(2
%
of reading
+
2 counts)
AC Real Range :
0.000mA – 30.00mA
Resolution: 0.001mA or
0.01mA
Accuracy:
±
(3% of r 0.05mA) All Ranges
PF > 0.1
V > 250VAC
DC Range 1:
0.0
µ
A – 350.0
µ
A for 7650 only
Resolution: 0.1
µ
A
Range 2:
0.300 mA – 3.500 mA for 7650 only
Resolution: 0.001mA
Range 3:
3.00 mA – 9.99 mA for 7650 only
Resolution: 0.01mA
Accuracy:
±
(2
%
of reading
+
2 counts)
DC Output Ripple
≤
4
%
Ripple RMS at 5 KV DC @ 10 mA, Resistive Load
Discharge Time
≤
200 ms
Summary of Contents for HYPOTULTRA III 7620
Page 7: ......
Page 8: ......
Page 130: ...122 configure it through the drive s properties in Windows ...