64
Abort
This message appears on the display if the test in process is aborted with the RESET button or
remote Reset control.
Max-Lmt
This message appears on the display if the DUT measurement exceeds the Max-Lmt setting of
any parameter.
Min-Lmt
This message appears on the display if the DUT measurement drops below the Min-Lmt.
CONT-Fail
This message appears on the display if the DUT fails the basic continuity check performed
during an AC/DC Withstand test (if Continuity is selected “ON”).
Arc-Fail
This message appears on the display if the DUT arcing current exceeds the Arc Sense limit and
Arc function is active (Arc Sense = 1… 9) of the AC/DC Withstand test.
Short
This message appears on the display if the DUT current is well beyond the metering range of the
test.
Charge-LO
This message appears on the display if the leakage current during Ramp-up falls below the
Charge -LO setting.
Breakdown
This message appears on the display if the DUT current is well beyond the metering range of the
test and the arcing condition beyond the arc sense limit.
GND-Fault
This message appears on the display if the GFI threshold is exceeded during the test.