65
Output Error
This message appears on the display, if the instruments output reading does not match the
setting. This message will only be seen if the EXIT key is press at the Output Error screen. If
the instrument has an output problem when the TEST button is pressed, the Output Error
screen will appear as follows:
O U T P U T E R R O R !
T h e i n s t r u m e n t h a s m a l f u n c t i o n e d !
S e t t i n g : x . x x k V
R e a d i n g : x . x x k V
P r e s s E X I T t o c o n t i n u e o r c a l l
1 - 8 0 0 - 8 5 8 - 8 3 7 8
The RESET button is not active in this situation. Only the EXIT key will allow you to return to
the Perform Test screen.
Fatal Error
If the instrument has a Fatal Error failure then the following screen will appear:
F A T A L E R R O R 9 0 0 X
T h e i n s t r u m e n t h a s m a l f u n c t i o n e d !
P l e a s e c a l l 1 - 8 0 0 - 8 5 8 - 8 3 7 8
All of the buttons and keys are not active in this situation. The customer should contact
Associated Research, Inc to receive further instruction.
Fatal Error identification number will represent type of the failure that occurs.
Error number of 9002
will appears on the display, if the instrument’s System data or
Model/Option/Serial Number data are corrupted and does not match the setting.
Error number of 9003
will appears on the display, if the instrument’s Calibration data is
corrupted.