93
If you set Arc Detection to ON, the Arc Sense is used to set the sensitivity for the Arc
Detection function. Use the numeric keypad to enter a value for the Arc Sense in the
specified range on the screen and use the enter key (
↵
)
to save the entry and move to
the next test parameter, Offset:
You can only set Tthe Offset value by performing the auto offset. To perform an auto
offset disconnect the DUT and leave all the test leads open. Press the test button on
the front panel. The instrument will display Reading on the screen when the
performing the auto offset. Once the instrument is finished performing the auto offset
the numeric value will be displayed on the screen and will be automatically saved to
be used when the actual test is performed. Use the Next arrow to move to the next
test parameter, Frequency: