151
scanners installed or connected to the instrument. Values 0
– 4 will indicate if there
are no scanners connected, one Internal or External scanner, or if both an Internal
and an External scanner are connected.
LF?
Lists the file name of the memory loaded into active memory (RAM).
LFN?
Lists the memory number of the active memory file loaded into active memory (RAM).
LF <memory number>?
List the file name of any of the 50 memories.
LP?
Lists the prompt that is created for the selected step within active memory (RAM).
LP <step number>?
Lists the prompt that has been created for a particular step of the file within active
memory (RAM).
LS?
Lists all the parameters for the individual step that is currently selected. See the ADD
command for the list of parameters. A comma (,) will separate each parameter and
will be preceded with the step number.
LS <step number>?
Lists all the parameters for the individual step indicated by
step number
= 1-30. See
the ADD command for the list of parameters. A comma (,) will separate each
parameter and will be preceded with the step number.
IEEE 488.2 Common Commands
These commands are required by the IEEE-488.2 standard with the exception of
*PSC, *PSC?. Most of these commands are not available over the USB/RS-232 bus
except for the *IDN? command which can be used to retrieve the instrument
identification information, and the four status reporting commands *ESR?, *ESE,
*ESE? and *STB?
COMMAND
NAME
DESCRIPTION
*IDN?
Identification Query
Associated Research Inc., Model
Number, Serial Number, Firmware
Revision
*RST
Reset Command
Resets HYPOTULTRA
*TST?
Self-Test Query
00H=OK
01H=TEST EEPROM ERROR