109
You can choose to set the Charge-Lo parameter automatically or manually. To
perform an auto Charge-Lo connect the DUT and all the test leads as needed to
perform a real test. Press the test button on the front panel. The instrument will display
Reading on the screen when performing the auto Charge-Lo. Once the instrument is
has finished performing the auto Charge-Lo it will display the numeric value on the
screen and will be automatically saved. The instrument will use this value when you
perform an actual test. Use the Next arrow to move to the next test parameter.
To enter a Charge-Lo value manually, use the numeric keypad to enter a value for the
Charge-Lo in the specified range on the screen and use the enter key (
↵
)
to save the
entry and move to the next test parameter, Int Scanner:
The Internal Scanner option is available on the HypotULTRA. See the Options section
for more information.
To set the scanner channels use the back (<) and forward (>) arrows to select
scanner channels and set the channels to High (H), Low (L) or Open (O). Use the
enter key (
↵
)
to save the values and move on to the next test parameter, Prompt: