181
COMMAND
NAME
DESCRIPTION
operations complete, Output
Queue=1
*WAI
Wait-to-Continue Command
*PSC {1|0}
Power-on Status Clear Command 1 = Power-on clear enable
registers
0 = Power-on load previous enable
registers
*PSC?
Power-on Status Clear Query
*ESR?
Standard Event Status Register
Query
0 - 255
*ESE <
value
> Standard Event Status Enable
Command
value
= 0 - 255
*ESE?
Standard Event Status Enable
Query
0 - 255
*STB?
Read Status Byte Query
Read Status Byte
*SRE
<
value
>
Service Request Enable Command
value
= 0 - 255
*SRE?
Service Request Enable Query
0 - 255
The following specifications apply to the Continuity Test with this option:
CONTINUITY TEST
Output Current, DC
0.1A for 0.00-
10.00Ω, 0.01A for 10.01-100.0Ω, 0.001A for 100.1-
1000Ω, 0.0001A for 1000-2000Ω.
All ranges have 10% over range capability.
Range indicated is for total resistance which includes any programmed
offset value.
Therefore when offset is used the meter display may fall into a higher
resistance range
SETTINGS
Max-
Lmt, Ω
0.00
– 9.99
0.01
± (1% of s 3 counts)
10.0
– 100.0
0.1
101
– 1000
1
1001
– 2000
1
± (1% of s 10 counts)
Resistance Display, Ω
0.00
– 9.99
0.01
± (1% of s 3 counts)
10.00
– 20.00
.01
± (1% of s 30 counts)
20.1
– 100.0
0.1
± (1% of s 3 counts)
100.1
– 199.9
± (1% of s 30 counts)
200
– 1000
1
± (1% of s 3 counts)
1001
– 2000
± (1% of s 10 counts)
Summary of Contents for HypotULTRA 7800
Page 34: ...29 ...
Page 53: ...48 ...
Page 121: ...116 ...
Page 128: ...123 ...
Page 143: ...138 ...
Page 187: ...182 ...
Page 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...