103
:
The Continuity function within a DCW test is a basic ground continuity check that can
be set to ON or OFF. Use the touchscreen keypad to set the Continuity function and
use the enter key (
↵
)
to save the entry and move to the next test parameter, Range:
You can set the Range to Auto or Fixed. Use the touchscreen keypad to set the
Range and use the enter key (
↵
)
to save the entry and move to the next test
parameter, Int Scanner: