4.12 Test Parameters
This section details the various test parameter descriptions. For information on setting
up test sequences, refer to section
5.
4.12.1 Description of Test Parameters
This section details each test parameter and test parameter description.
Voltage
: The voltage that is applied to the high voltage and return terminals during a
test.
HI-Limit T
: ACW test only. Total current maximum limit. The maximum current
threshold that triggers a failure when exceeded.
HI-Limit R
: ACW test only. Real current maximum limit. The maximum current or
resistance threshold that triggers a failure when exceeded.
LO-Limit T
: ACW test only. Total current minimum limit. The minimum current or
resistance threshold that triggers a failure when not exceeded.
LO-Limit R
: ACW test only. Real current minimum limit. The minimum current or
resistance threshold that triggers a failure when not exceeded.
HI-Limit
: The maximum current or resistance threshold that triggers a failure when
exceeded.
LO-Limit
: The minimum current or resistance threshold that triggers a failure when
not exceeded.
Ramp Up
: The length of time that is allowed for the test voltage to climb from 0 to the
programmed test voltage.
Dwell
: The length of time that is allowed for the programmed test voltage to be
applied.
Delay
: The length of time that the programmed test voltage is applied but no judgment
of the set parameters is made. Judgment of the parameters is not made until the end
of the delay time.
Ramp Down
: The length of time that is allowed for the test voltage to decay from
programmed test voltage to 0.
Arc Sense
: The maximum allowable threshold for arcing. The numbers 0 through 9
correspond to the different arc sensitivity levels, 1 meaning the maximum threshold of
allowable arcing, 9 meaning the minimum threshold of allowable arcing, and 0 being
OFF. Arc detection is not required for testing.