35
When the PLC remote is turned ON, the front panel TEST button is disabled and a
you can only start a test through the rear panel I/O. If you attempt to start a test from
the front panel TEST button when the PLC Remote function is turned “ON”,
HypotULTRA displays a pop-up message. From the main Setup System menu touch
the PLC Remote icon and the following screen will appear:
4.4.3 Single Step
Use this function to temporarily override the automatic connection feature. When the
Single Step function is ON the instrument will pause after each step is completed. To
continue the test sequence, press the Test button to execute the next connected step.
Each time the TEST button is pressed the next connected step will execute. If you
press the RESET button before completing all connected steps, it will return you to the
original starting step. If a step fails and you wish to continue to the next step, do not
press RESET.
From the Setup System menu touch the Single Step icon and the following screen will