149
COMMAND
NAME
VALUE
Level: 0=Run Only,
1=Recall Setup,
2=Edit Step,
3=Full System
SUD n
User Delete
n=1-5
SUE p,p,p,p
User Edit
SUA
Pointer,Name,Password,Level
Pointer: 1-5
Name: 8 character max
Password: 8 number max
Level: 0=Run Only,
1=Recall Setup,
2=Edit Step,
3=Full System
SM n
SM?
Measurement
n=0-1
0=True RMS, 1=Average
SBI n
SBI?
Barcode IP Format
n=0-4
0=Off, 1=Ser/Prod, 2=Serial#,
3=Product#, 4=Run File
SAS n
SAS?
Auto Start Format
n=0-1
0=Off, 1=On
Query Commands
These query commands will retrieve data from the instrument. The GPIB bus
application requires an IEEE-488 read command to be sent after the query command.
These commands include functions for retrieving test data, test results and remote
hardware status, as well as setup file information.
COMMAND
NAME
VALUE
TD?
List Testing Data
Test In Process
RDM?
Read Max Current
ACW: Total current,
Real Current DCW:
current
RD <
step
number
>?
List Results Data
step number
= 1-200
RR?
Read Remote Reset
1=Open, 0=Closed
RI?
Read Remote Interlock
1=Open, 0=Closed
RS?
Read Scanner Status
0=None, 1=Int, 2=Ext
1, 3=Int+Ext 1,
4=Ext 2, 5=Int+Ext 2,
6=Ext 1+Ext 2,
7=Int+Ext 1+Ext 2
RU?
Read USB Stick
0=NO, 1=YES
Summary of Contents for HypotULTRA 7800
Page 34: ...29 ...
Page 53: ...48 ...
Page 121: ...116 ...
Page 128: ...123 ...
Page 143: ...138 ...
Page 187: ...182 ...
Page 202: ...197 Calibration of GFI 5 0mA The calibration of GFI 5 0mA is a two step process ...