161
The three modes available are Auto, Manual and OFF. In the Auto mode, the
HypotULTRA will automatically send the test results to the printer at the end of
every complete test. If Manual mode is selected, at the end of the test you will
need to select the Results icon to view the results. Under Results you will then
select Present test. The next screen will display the results of the present test
along with an option of Print Results. You can then select Print Results to send
the test results to a printer.
OFF mode disables the printer card and printer menu.
Test Result
From the Print Format Setting screen, select “Test Result”:
The two options available for Test Result are “All” and “Fail”. This option allows
you to select which test results will b
e printed. “All”, selects printing all test results
and “Fail only”, selects printing only failed test results.