ACW 400/800Hz Calibration (optional):
If you have option 13 (ACW 400/800Hz)
installed, there are two additional calibration selections (ACW VOLT-400Hz and AC
VOLT-800Hz):
10.3. Calibration of Parameters
From the Calibration screens, use the touchscreen to select the parameter you wish to
calibrate. A calibration prompt screen will now appear that describes the necessary
load and connection information for the parameter being calibrated.
Once you press TEST, the Calibration data entry screen will appear for the selected
parameter. Read the measurement from your standard and enter it using the numeric
keypad. Once a value is entered the screen for the next step will appear.
Calibration of AC Hipot Voltage
The calibration of AC Hipot Voltage is a two-step process. The instructions for the first
step appear as follows: