33
Measure
Every
\HDU
cycles
Every other
\HDU
cycles
Time*
(minutes)
3
Process tank
3.1 Clean as required.
•
-
3.2 Check and clean the level sensor.
•
5
3.3 Check that the valves are working and that the
connections are not leaking. Clean as required.
•
10
4
Steam Generators
4.1 Check the connections to the steam generator
for leakage and make sure the surrounding
insulation is intact and no hot surfaces are
exposed.
•
5
4.2 Check that the steam generator is working
properly.
•
10
5
Main pump
5.1
&KHFNWKDWWKHPDLQSXPSLV¿UPO\DWWDFKHGDQG
that there are no leaks in the connections.
•
5
6
Dosing system
6.1 Check the hoses between the dosing pump and
the detergent holder. Replace if necessary.
•
5–10
6.2 Check the dosage amount. Adjust if necessary.
•
10–30
6.3 Replace hose in the dosing pump.
•
5
6.4 Check that the empty container alarm works.
Clean or replace as necessary.
•
5
6.5 Check the function of the check valve in the
steam generator.
•
5
7
Drainage connection
7.1
&KHFNWKDWWKHGUDLQDJHFRQQHFWLRQLV¿UPO\
attached and that it is not leaking.
•
5
8
Ventilator (Option)
8.1
(3$¿OWHU
•
10
8.2 Check the fan for proper operation and check
that the connections are tight.
•
5
8.3 Check that the check valves are functioning
properly. Clean as required.
•
10
8.4 Check hoses to the fan.
•
5
10
Function check (at each service)
10.1 Run a complete process and be sure there are no deviations according to the
periodic checkpoints.
10.2 Check that the panel’s lights and alarms are functioning.
* The time required is estimated with the top off and may vary depending on the installation settings,
operating conditions and equipment level.