23
([WHUQDOFRROLQJOLQH
The line is used to set the operating time for the ventilator.
• 0–600 seconds in one-second steps
• Default value is 120
Changing the settings requires authorization level B or C.
4.8 Disinfection
'LVLQIHFWLRQWHPSHUDWXUHOLQH
The line is used to set the temperature at which the disinfection
time in line 16 is to begin.
• 80–95°C (can only be set in °C)
• 80°C = 176°F
• 85°C = 185°F
• 90°C = 194°F
• 95°C = 203°F
• Default value is 91°C
Changing the settings requires authorization level C.
'LVLQIHFWLRQWLPHOLQH
The line is used to set the time for thermal disinfection.
• 0–1800 seconds in one-second steps
• Default value is 60
Changing the settings requires authorization level C.
3UHKHDWLQJVWHDPJHQHUDWRUOLQH
The line is used to set the time for preheating the steam generator.
• 0–90
seconds
• 0 = Shut down
• Default value is 50
Changing the settings requires authorization level C.
&RROLQJWLPHOLQH
The line is used to set the operating time for the ventilator after the
process.
• 0-600
seconds
• Default value is 120
Changing the settings requires authorization level C.