Note!
)RUFHUWDLQRXWSXWVWREHDFWLYDWHGWKHGRRUPXVWEHFORVHG
and locked.
The lines are used to control the digital outputs. Authorization level
B or C required.
Line
Output
Function
61
01
Valve, cold water
62
02
Valve, hot water.
63
03
Ventilator
64
04
Pump, rinse-aid
65
05
Pump, descaler
66
06
Main pump
67
07
Steam Generators
68
08
Pump, process agent
69
09
Unlock door (1 runs unlocking
sequence)
70
10
Lock door (1 runs locking sequence)
First press
or
to select either:
• 00 = Output not active
• 01 = Output active
Then press
to control the output.
5HDGLQJWHPSHUDWXUHOLQH
The line shows the current temperature sensor value in the chosen
temperature scale. This value cannot be changed.
Display range: 1–100°C/34–212°F in 1° intervals.
5HDGLQJLQGHSHQGHQWWHPSHUDWXUHOLQH
The line shows the current temperature sensor value in the chosen
temperature scale. This value cannot be changed.
Display range: 1–100°C/34–212°F in 1° intervals.
&DOLEUDWLQJWHPSHUDWXUHVHQVRUVOLQHV±
(this functionality is not available for AH1036512)
Used to calibrate temperature sensors.
• Line 77 is used for external temperature calibration.
• Line 78 is used for external temperature calibration.