28
5 ERROR
INDICATORS
5.1 Error
message
For certain errors in the process, information about the cause
of the error is given on the display on the front panel. The error
messages must be acknowledged when the error has been put
right. The following error messages may appear:
5.1.1 Error codes (must be acknowledged):
Line
Function
Comments
F1
Temperature sensor error
Temperature outside interval 1-130 or
internal/external difference >3°C.
F4
Door logic error
Door locked but not closed.
F7
Disinfection temperature not reached
Temperature not reached within 900s.
F8
Door open/unlocked during process
The door opened during the process.
F9
Checksum error
E2 memory error. Replace CPU card.
F10
&RXOGQRW¿OOHPSW\WDQN
1RW¿OOHGDIWHUVRUHPSWLHGDIWHUULQVLQJ
F11
Leakage
/HDNDJHVHQVRULQRYHUÀRZLVDFWLYHRUWDQN
level active in standby. Leakage sensor is an
RSWLRQRQWKHÀRRU
F13
Internal software error
Reset the machine
5.1.2 Information codes (acknowledgement not needed):
Warning Meaning
Measure
U1
Lime binding agent low
Top up lime binding agent
U2
Process agent low level
Top up process agent
U3
Rinse-aid low level
Top up rinse-aid
U7
Warning of high goods temperature
Let the goods cool down
U8
External communication
U9
Inactivity warning
The machine has been inactive for
72 hours. A program with an empty
chamber must be run before using
the machine to clean goods.
5.1.3 Handling codes that require action before restarting the machine:
Warning Meaning
H6
Start attempt with door open
H7
Cannot lock door
H8
Cannot unlock door