3-10 Pulse Modulation Tests
Chapter 3 — Performance Verification
3-42
PN: 10370-10372 Rev. B
MG3702xA MM
c.
Press the
SPAN
key and enter 100 kHz.
d.
Press the
MKR
key, then select
MARKER DELTA
to set the marker reference.
3.
On the MG3702xA's External Pulse Status menu, select On/Off to turn the pulse mode on.
4.
On the Spectrum Analyzer, read the marker delta value and record the value as P
depth
in the test
record.
5.
for each frequency listed in the test record.
Note
The signal level may drift slowly after the pulse mode is turned on. (Make the measurement as soon
as possible after turning the pulse mode on.) This drift is the result of the Sample/Hold circuit not
holding the level because of a very low pulse duty factor (time ratio of RF ON to RF OFF). This drift
will not be present in normal pulse operation as the minimum pulse repetition rate is 100 Hz.