![Anritsu MG3702xA Series Maintenance Manual Download Page 43](http://html.mh-extra.com/html/anritsu/mg3702xa-series/mg3702xa-series_maintenance-manual_2945427043.webp)
MG3702xA MM
PN: 10370-10372 Rev. B
3-1
Chapter 3 — Performance Verification
3-1
Introduction
This chapter contains tests that can be used to verify the performance of the series MG3702xA Synthesized
Signal Generators to specifications. These tests support all instrument models having any version of firmware
and instrument models with the following options:
•
Option 001 (rack mount with slides)
•
Option 002 (mechanical step attenuator)
•
Option 004 (digital down converter)
•
Option 009 (rear panel RF output which moves the RF output to the rear of unit)
•
Option 011 (rack Mount without slides)
•
Option 015 (high power output)
•
Option 017 (deletes the front panel)
•
Option 026 (pulse modulation)
•
Option 035 (removable hard drive)
•
Option 036 (ultra stable phase track)
•
Option 037 (performance suite) see options 002, 004, 015 and 026
•
Option 088 (250 µsec Switching speed limit for export control)
3-2
Test Records
A blank copy of a sample performance verification test record for the MG3702xA models is provided in
Appendix A. Each test record contains the model-specific variables called for by the test procedures. It also
provides a means for maintaining an accurate and complete record of instrument performance. We recommend
that you copy these pages and use them to record the results of your initial testing of the instrument. These
initial test results can later be used as benchmark values for future tests of the same instrument.
3-3
Connector and Key Notation
The test procedures include many references to equipment interconnections and control settings. For all
MG3702xA references, specific labels are used to denote the appropriate menu key, data entry key, data entry
control, or connector (such as RF Output). Most references to supporting test equipment use general labels for
commonly used controls and connections (such as Span or RF Input). In some cases, a specific label is used that
is a particular feature of the test equipment listed in
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......