Chapter 3 — Performance Verification
3-9 Power Level Accuracy and Flatness Tests
MG3702xA MM
PN: 10370-10372 Rev. B
3-25
a.
Preset the instrument by pressing the green
PRESET
button.
After preset, the CW menu is displayed.
Test Procedure
1.
Set up the MG3702xA as follows:
a.
Press
Frequency
to open the current frequency parameter for editing.
b.
Press
Edit Frequency
and set F0 to the CW frequency indicated in the test record.
c.
Press
Edit Level
to open the current power level parameter for editing.
d.
Set L0 to the first applicable power level indicated in the test record.
2.
Measure the output power level with the power meter and record the reading in the test record.
3.
On the MG3702xA, use the cursor control key (Arrow keys) to decrement L0 to the next test power
level in the test record. Measure and record the power meter reading in the test record.
4.
Repeat
for each of the test power levels listed in the test record (down to –50 dBm) for the
current CW frequency.
5.
Repeat
for all CW frequencies listed in the test record.
Power Level Accuracy (Models with Option 002)
For the following test, refer to the test record table in
,
“Section 3-9 Power Level Accuracy and
“Power Level Accuracy (Models with Option 002)”
Power level accuracy for power levels below –50 dBm is tested in two methods:
•
Measuring the MG3702xA’s RF output directly on a measuring receiver
•
Down converting the MG3702xA’s RF output and measuring the down converted IF on a measuring
receiver
In both cases, a reference power level is set on the measuring receiver and the output power level is stepped
down in 5 dB increments. The relative output power level is then measured at each step.
Caution
To prevent damage to the power sensor, use a fixed attenuator when measuring power levels above
+19 dBm.
Figure 3-10.
Equipment Setup for Power Level Accuracy and Flatness Tests Below –50 dBm
MG3702xA (DUT)
Local Oscillator
Measuring Receiver
10 MHz
REF OUT
10 MHz
REF IN
Mixer
Down Conversion Path
Through Path
RF
Input
10 MHz
Reference In
Connect the dashed lines
as directed by the procedure.
Note: Make this connection as
short as possible by using an
M to M adapter.
MG3694B
Synthesizer
10 MHz to 40 GHz
MG37022A
Signal
Generator
Standby
Operate
ML2530A
ML2530A
Calibration Receiver
L
I
R
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......