3-7 Spurious Signals Tests
Chapter 3 — Performance Verification
3-10
PN: 10370-10372 Rev. B
MG3702xA MM
b.
Press
REF LVL
and enter the current power level setting (L1) of the MG3702xA.
c.
Press the
FREQUENCY
key.
d.
Press
Center Freq
.
e.
Enter the current spectrum analyzer center frequency from the test record.
f.
Press
SPAN
.
g.
Enter the current spectrum analyzer span frequency from the test record.
h.
Press the
BW
key.
i.
Press
Res BW to MAN
.
j.
Enter the current spectrum analyzer RBW value from the test record.
3.
Press the
PEAK SEARCH
key, and then select
MARKER DELTA
, which sets the reference for the
following measurements.
4.
Set up the spectrum analyzer as follows:
a.
Press the
FREQUENCY
key.
b.
Press
Center Freq
.
c.
Enter the next spectrum analyzer center frequency from the test record.
d.
Press the
BW
key.
e.
Press
Res BW to MAN
.
f.
Enter the next spectrum analyzer RBW value from the test record.
5.
Press
PEAK SEARCH
.
6.
Record the reading from the spectrum analyzer Delta MKR reading in
,
Non-Harmonic Signals Test (Part 1)”
“Section 3-7 Non-Harmonic Signals Test (Part 2)”
Part 2.
7.
Repeat
through
for same specified MG3702xA frequency listed in the test record,
part 2.
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......