Chapter 3 — Performance Verification
3-9 Power Level Accuracy and Flatness Tests
MG3702xA MM
PN: 10370-10372 Rev. B
3-33
b.
Press
Edit F2
, enter the upper frequency of the MG3702xA under test using the next
frequency range value in the test record.
c.
Press
Dwell time
and set to 250ms.
d.
Press
Num of Steps
to open the number-of-steps parameter for editing.
e.
Enter the number of steps called out in the test records.
Test Procedure
1.
On the MG3702xA, press
Trigger
in the
Step Sweep
menu to start the sweep.
2.
Monitor the power level and the DMM.
a.
On the power meter determine the Minimum power values.
b.
Monitor the DMM which will be incrementing from approximately 0v to approximately 10V.
When the DMM has reached approximately 10V the sweep is complete.
c.
Repeat
and
until you have the Minimum power reading.
d.
Add the attenuation value of the attenuator to the minimum power reading and record this
value to the test record. Verify that the reading is above the specification noted in the test
record.
3.
Setup the power meter for the next measurement.
a.
Configure the power sensor's calibration factor source by pressing:
Sensor
|
CalFactor | SOURCE
until V/GHZ is displayed.
b.
Setup the minimum V/GHz range by pressing:
Setup
|
Start F
c.
Enter the lower frequency of the MG3702xA under test using the next frequency range value
in the test record.
d.
Setup the maximum V/GHz range by pressing:
Setup
|
Stop F
e.
Enter the upper frequency of the MG3702xA under test using the next frequency range
value in the test record.
f.
Press any hard key to begin the measurement.
4.
Repeat the setup and test starting at
of
until all tests are
completed in the test record.
Note
Be sure to use and calibrate the appropriate power sensor for the frequency being measured.
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......