UG-1828
Preliminary Technical Data
Rev. PrC | Page 292 of 338
Figure 270. ADRV9001 Evaluation Card and ZYNQ ZC706 Evaluation Platform with Connections Required for Testing
To set up the evaluation board for testing, follow steps listed below:
1.
Connect the ADRV9001 evaluation card and the ZYNQ ZC706 evaluation platform together as shown in Figure 270. Use the LPC
FMC connector (J5). Take care to be sure the connectors are properly aligned.
2.
Make sure that all jumpers on the ZYNQ ZC706 evaluation platform as well as the SW11 position (1, 2, 5 = “A” position) match
settings shown in Figure 269.
3.
Insert the SD card that came with the ADRV9001 evaluation kit into ZYNQ ZC706 evaluation platform SD card slot (J30).
4.
On the ADRV9001 evaluation card, provide a device clock (frequency must match the setting selected in the TES), at a +13dBm
power level to J501 connector. (This signal drives the reference clock into the ADCLK944 clock distribution chip on the board – the
Q1/Q1_N pins of ADCLK944 generates the DEV_CLK for the ADRV9001 and REF_CLK for the Xilinx FPGA on the ZYNQ
platform).
a.
It should be noted that quality of clock source used to generate DEV_CLK will directly impact overall system performance. User
must ensure that high quality, stable and low phase noise clock source is used here.
5.
Connect a 12V, 5A power supply to the ZYNQ evaluation platform at the J22 header.
6.
Connect the ZYNQ evaluation platform to the PC with an Ethernet cable (connect to P3). There is no driver installation required.
a.
In the case when the Ethernet port is already occupied by another connection, use an USB-to-Ethernet adapter.
b.
On an Ethernet connection dedicated to the ZYNQ platform, the user must manually set the following:
i.
IPv4 Address to: 192.168.1.2
ii.
IPv4 Subnet Mask to: 255.255.255.0
Refer to Figure 271 for more details.
The user should make sure that ports listed below are not blocked by firewall software on their PC:
•
22—SSH protocol
•
55557—access to the evaluation software on ZYNQ platform
Note that the ZYNQ ZC706 evaluation platform IP address is set by default to: 192.168.1.10.
ETHERNET
CONNECTION
PC RUNNING TRANSCEIVER
EVALUATION SOFTWARE
50Ω
TERMINATION
SD CARD FROM
ADRV9001 EVALUATION KIT
SWITCHING
POWER SUPPLY
(12V DC)
POWER
SWITCH
J501
Tx1
Tx2
Rx2A
Rx1A
EXTERNAL
LO 1
SIGNAL GENERATOR
Rx2B
Rx1B
SIGNAL GENERATOR
SIGNAL ANALYZER
EXTERNAL
LO 2
24159-