3) Fit the new aperture and grating, and replace the check reflector.
4) Repeat the clear-path calibration and check the B value. If it is within
the range 0.95 to 1.05, the B value is acceptable. Otherwise repeat the
procedure.
Example
The B-value is 1.23, Grating 2, Aperture F
•
Sequence number is 25.
•
B-value is 0.20 too high.
•
0.20 = 4 x 0.05, so the required sequence number is 29
•
Refer to Section F6. The sequence number is 29. Therefore select
Grating 2, Aperture J.
F4 Calibration Error Test
Perform a calibration error test according to ASTM D6216-12 para 7.8. Select
three audit filters selected according to the requirements of ASTM D6216
para 7.5. The 2” square audit filters provided as an optional item with the
Model 4500 MkIII are suitable. Ensure the calibration certificates are up-to-
date.
1) Open the 4500 MkIII transceiver at the hinge and place the low-level
filter in the audit filter slot.
2) Close the transceiver and wait for the reading to settle. Record the
opacity value indicated on the site data acquisition system. As a
precaution, it is advisable to write down the value displayed on the
Model 4500 MkIII user interface.
3) Repeat steps 1 and 2 for the mid- and high-level opacity filters.
4) Repeat steps 1, 2 and 3 four more times, to give a total of five readings
for each filter.
5) Calculate the calibration error as described in D6216 para 7.8.
Fig. F-1
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...