Check the software and replace the instrument cover.
1) Cycle the instrument power Off then On.
2) Ensure that after a few seconds, the calibration motors move, the
display shows the startup splash, followed by the version numbers of
the Controller and Human Interface firmware.
3) Check that the version numbers are those expected from the upgrade.
4) Remove the power.
5) Carefully replace the cover, ensuring that it doesn’t jam on any
components. When the cover is close enough, refit the keyboard
connector ribbon, ensuring it is not offset by one pin.
6) Ensure that the white cord seal is correctly seated before pushing the
cover home. Secure the cover with the eight screws.
7) Switch on the instrument and check display and keyboard functionality.
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...