![Ametek Land 4500 MKIII User Manual Download Page 51](http://html1.mh-extra.com/html/ametek-land/4500-mkiii/4500-mkiii_user-manual_2931437051.webp)
B - 9
Model 4500 MkIII
Opacity and Dust Monitor
User Guide
Note
If any of the settings are changed from their default values, the new
values should be noted on the Calibration Report and Settings List
supplied with the instrument.
B4.5 Setup Calibration Check Interval
1)
From the instrument reading display,
press
2) Unlock
is displayed, press
3) Use the and arrows until the Supervisor code (Default: 10) is
displayed, wait for the
icon to appear and press
4) Use the and arrows to select Settings
and press
5) Use the and arrows to select Cal Check Interval
and press
6) Use the and
arrows to change Cal Check Interval (in hours), wait
for the icon to appear and press
7) Press to return to Settings
Note: Auxiliary Functions Unit (AFU)
If an AFU is fitted, the Transceiver outputs are not used, and there is no
need to perform the settings described in sections 6, 7 and 8 below. AFU
output setting instructions are given in sections 10, 11 and 12.
B4.6 Setup 4-20 mA Current Loop Output
The Model 4500 MkIII normally provides the power supply for the 4-20 mA
current loop output. It is intended to be connected to a passive input device.
To connect the 4-20 mA current loop output to an active device, option links
inside the Transceiver must be changed. Switch off the power. Remove the
rear cover from the Transceiver (see section E3) and locate Link 1 and Link 2
on the Main printed circuit board.
The Model 4500 MkIII is shipped set for Active output with the Links in the
following positions:
Link 1:
A
Link 2:
B
For Passive output, set the Links in the following positions:
Link 1:
B
Link 2:
A
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...