A - 1
Model 4500 MkIII
Opacity and Dust Monitor
User Guide
A
Introduction
A1 General Description
The Land Instruments International Model 4500 MkIII Continuous Opacity
Monitoring System (COMS) measures opacity by shining a light beam
through flue gases.
An internal microprocessor calculates dust density and other parameters.
The Model 4500 MkIII is designed for continuous operation in all weather
conditions. Maintenance requirements are minimal.
Transceiver
Containing all of the major electronic and electro–optic
components.
Retro-Reflector
Containing a corner cube reflector.
Air Purge System A continuous supply of purge air is essential to prevent
dust and corrosive gases from affecting the optical
system. Single and dual electric blowers or compressed-
air driven devices are available to suit individual site
requirements. Automatic fail–safe shutters can also be
fitted for temporary protection in the event of a purge air
failure.
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...