![Ametek Land 4500 MKIII User Manual Download Page 87](http://html1.mh-extra.com/html/ametek-land/4500-mkiii/4500-mkiii_user-manual_2931437087.webp)
D - 1
Model 4500 MkIII
Opacity and Dust Monitor
User Guide
D
System Specification
D1 4500 MkIII Dust and Opacity Monitor
Measuring
Technique:
Double pass transmissometry
Operating Wavelength:
520 ± 20 nm
Light Source:
Pulse High Intensity LED
Range:
Opacity 0 – 10 % to 0 – 100 %
Optical Density 0 – 0.1 to 0 – 3.0
Dust Density 0-10 to 0 to 10000 mg/m
3
User
Selectable
Calibration Error:
1.5 % Opacity
Drift:
<0.3 % Opacity per month
Thermal Stability:
0.6 % Opacity / 22 °C change
Angle of Projection:
2
o
Angle of View:
2
o
Response Time:
<10 seconds to 95 % of final value
Averaging:
Selectable from 10 seconds to 24 hours (in 1 sec steps)
Pathlength:
0.5 to 10 m / 20 in to 32 ft (extended pathlengths
to 15 m / 50 ft available
Calibration:
Automatic zero and upscale check
(Selectable interval 1 to 24 hours in 1 hour steps)
Zero Correction:
Automatic correction for zero drift
Control Panel
Display:
128 x 64 pixel reflective backlit LCD
Keypad:
4 keys for data input
Indicators:
Power, System OK, Calibration, Alarm
Environmental
Operating temperature:
-40 to 55
o
C / -40 to 131
o
F
Maximum Flue Gas temp:
600
o
C / 1112
o
F
Maximum Flange temp:
200
o
C / 392
o
F
Flue Gas Water Content
Not sensitive to water vapor. Condensed
water must not be present in the
measurement path
Electrical Safety:
IEC / EN 61010-2
EMC:
EN50 081 & EN50 082
Sealing:
IP65 / NEMA4X
Approvals:
EN15267-1, EN15267-2, EN15267-3,
EN13284-2
Performance Specifications:
QAL1, US EPA PS-1, ASTM D6216,
US
EPA
Procedure
3
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...