A4.4 Connecting the Transceiver to the AFU
Only connectors A and C are needed to connect the Model 4500 MkIII transceiver
to the AFU. All connections are made to the AFU main circuit board, not the
daughter board. Customer connections can then be made directly to the AFU.
Connector A
Function
Colour
AFU Terminal
24 V dc
Red
32
0 V dc
Blue
33
Screen
Braided Screen
34
Connector B (Optional)
This connection is required only if a third 4 to 20 mA signal will be used.
Function
Colour
AFU Terminal
4 to 20 mA +
Yellow
61
4 to 20 mA -
Green
62
Screen
Braided Screen
63
Connector C
Function
Colour
AFU Terminal
Screen
Braided Screen
55
Modbus D1
Green
56
Modbus D0
Yellow
57
Modbus Common
Black
58
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...