B - 27
Model 4500 MkIII
Opacity and Dust Monitor
User Guide
B5.6 Restoring Previous Calibration Values
If the Model 4500 MkIII is mis-calibrated, it is possible to recover the
situation by replacing the erroneous calibration constants. The Previous
constants, and the constants determined during Factory testing are stored in
the instrument.
To replace the Current Calibration constants with the Previous constants:
1) Press
2) Unlock
is displayed, press
3) Use the and arrows until the Supervisor code (Default: 10) is
displayed and press
4) Use the and arrows to select Calibration
and press
5) Use the and arrows to select Previous Cal
and press
6) Use the and arrows to select Restore Last Cal
and press
7) Confirm
is displayed. Press
8) The Previous Calibration constants are copied into the Current
Calibration constants.
9) The Tick
in the top left corner of the display indicates that the
operation was successful.
10) Press
three times to return to the main display.
To replace the Current Calibration constants with the Factory constants:
1) Press
2)
Unlock
is displayed, press
3) Use the and arrows until the Supervisor code (Default: 10) is
displayed and press
4) Use the and arrows to select Calibration
and press
5) Use the and arrows to select Previous Cal
and press
6) Use the and arrows to select Restore Factory Cal
and press
7) Confirm
is displayed. Press
The Factory Calibration constants are copied into the Current Calibration
constants.
8) The Tick
in the top left corner of the display indicates that the
operation was successful.
9) Press
three times to return to the main display.
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...