E1.1 Troubleshooting
Problem seen
Possible cause
Recommended solution
Transceiver display is
blank
No power
Check power supply
Check power cable
Hardware fault
Contact AMETEK Land for
advice
Transceiver does not
recognise AFU
Communications problem.
This can be identified by
looking at communications
LED23 (TXD) and LED24
(RXD) which should flicker
rapidly if communications
are normal
Check wiring
Check jumpers LK1 And
LK2 are fitted on AFU
AFU Switch settings
Check status of AFU
switches S2, S5, S7
AFU 4-20 mA outputs do
not function
Transceiver is not
communicating with AFU
As above
Unit does not output
opacity after calibration
Output is PLCF
No action needed.
Instrument is designed
to output the PLCF value
for 90 s at the end of a
calibration check
Unexpected calibration
checks
Internal timer is in use
Set parameter 95 to value
0 to disable calibration
check timer
Intermittent loss of power
causes instrument to
restart
Check power connections
are secure
Check mains power to
AFU is reliable
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...