Fault
Number
Description
Recommended Action
1
Zero motor jammed.
Check the Zero Reflector drive mechanism. Warning – Do not move the
mechanism by hand or the precision gearbox may be irreparably damaged.
2
Upscale motor jammed.
Check the Upscale Filter drive mechanism. Warning – Do not move the
mechanism by hand or the precision gearbox may be irreparably damaged.
3
Source LED failed
Check the Source LED connections. Replace the Source LED Assembly.
4
Flood LED failed
Check the Flood LED connections. Replace the Flood LED Assembly.
5
ADC Over-range
The signal arriving at the Analogue-to-Digital Converter is too large. Check
that the correct reflector element and aperture plate are fitted in the retro-
reflector. Replace the Detector PCB. Warning – The Detector PCB must be
optically aligned.
6
ADC Fault
The Analogue-to-Digital Converter is not working. Replace the Detector
PCB. Warning – The Detector PCB must be optically aligned.
7
Transceiver shutter fault
If fitted, the fail-safe shutter should close during calibration check and
then reopen.
Check shutter wiring.
Check shutter operation during calibration check.
Replace shutter motor or clutch.
Replace shutter. As a temporary measure, a faulty shutter can be locked
open to allow opacity measurements to continue. Refer to E3 below.
8
Retro shutter fault
As 7 above.
9
Transceiver blower fault
Check blower power.
Check pressure switch wiring.
Check blower filter.
10
Retro blower fault
As 9 above
11
AFU Fault
Communications lost to Auxiliary Functions Unit.
Check wiring to AFU. Check links LK1 and LK2 are fitted (refer to Section
A4.8).
Possible hardware fault in AFU - replace main board.
13
Negative Opacity
Check that the installation pathlength is correct. Perform a Clear Stack
Calibration.
14
Checksum Error
The microprocessor’s program memory is defective. Re-program the Main
PCB. Change the Main PCB. Warning – All of the instrument settings and
calibration constants are stored on the Main PCB. Re-enter this data after
changing the PCB.
15
Lens Contamination Limit
The zero drift correction system has reached its limit. Clean the main lens.
Perform a Calibration Check. If the problem persists, make sure the Zero
Reflector is clean. Note that this fault can only be cleared by a successful
Calibration Check or Clear Stack Calibration.
16
Failure during Calibration
Something has prevented completion of a calibration. User actions can
cause this fault. Perform a Calibration Check. If the problem persists,
check the Zero Reflector and Upscale Filter drive mechanisms. Warning –
Do not move either mechanism by hand or the precision gearbox may be
damaged irreparably.
Note
When contacting Land Instruments Service Department please have the
Instrument Type and Serial Number available. This will ensure a more
accurate and efficient response.
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...