Opacity and Dust Monitor
User Guide
Model 4500 MkIII
G - 2
G2 Model 4500 MkIII Parameter Record Sheet
The following parameters give valuable information on the operation of the Model
4500 MkIII. Recording them regularly (e.g. as part of a quarterly calibration audit)
can be valuable in diagnosing fault conditions.
Parameter Label
Description
Value
6
Software
Version number
7
HI Software
Version number
51
Faults
No. of faults
52
Error Flags 1
Fault code
121
Sm
Signal measurement
122
Rm
Reference measurement
123
Sf
Signal flood value
124
Rf
Reference flood value
125
So
Signal offset
126
Ro
Reference offset
127
Opacity
Calculated opacity
137
PLCF
Path length correction factor
140
C working
Calibration slope
141
X working
Calibration offset
142
B working
Calibration balance factor
149
W
Dirty window correction
153
Op0
Zero compensation
154
Opz
Cal check zero value
155
Opc
Cal check upscale value
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...