Opacity and Dust Monitor
User Guide
Model 4500 MkIII
H - 2
The unit can be accessed via a web browser, providing the static address is
within the allowed range on your network.
If it is not possible to connect the module directly onto your network using
this static address, then it will be necessary to setup a direct connection
between the module and a PC in order to configure the module. The following
describes how to do this using a PC running Windows XP with SP2.
Direct Connection
In order to configure the AFU Ethernet module, a direct connection to the
device is recommended. The following instructions provide the preferred set
up procedure: if in doubt contact your network Administrators.
1) Connect the cross-over Ethernet cable between the AFU & PC/Laptop.
2) Disable the ‘Proxy Server’ from the system, allowing direct IP
addressing (Internet Protocol - TCP/IP) between the laptop/PC and the
AFU Module.
3) The following instructions detail how to set the IP address of the AFU
instrument (static addressing) from within a Windows XP Operating
System and how to disable the ‘Proxy Server’ from ‘Internet Explorer’.
•
From the Windows XP desktop, click
Start
and
Settings
.
• Choose
Control Panel
.
•
Double-click on
Network Connections
.
•
Identify the network adapter in use (typically
Local Area
Connection
).
•
Highlight and right-click on
Local Area Connection
.
• Choose
Properties
.
• Highlight
Internet Protocol (TCP/IP)
.
• Click
Properties
.
• Select
Use the following IP address
.
•
Enter the IP address given with the AFU interface unit and add 1 to
the last number of the IP address:
i.e. 10.1.10.200 (AFU IP address)
10.1.10.201 (IP address to be entered)
•
Set the
Subnet Mask
to 255.255.0.0 (default) or identical to the
AFU and ensure that the
Default Gateway
is blank.
• Click
OK
(twice).
• Close
Control Panel
. The settings are applied automatically.
Summary of Contents for 4500 MKIII
Page 4: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 6: ...INTRODUCTION A...
Page 29: ...A 23 Model 4500 MkIII Opacity and Dust Monitor User Guide A4 3 Mounting the AFU APS I O...
Page 41: ...GETTING STARTED B...
Page 42: ...Opacity and Dust Monitor User Guide Model 4500 MkIII Preface Blank...
Page 76: ...THEORY OF OPERATION AND APPLICATION C...
Page 86: ...SYSTEM SPECIFICATION D...
Page 91: ...MAINTENANCE E...
Page 108: ...CHANGING THE INSTALLATION PATHLENGTH F...
Page 113: ...INSTRUMENT RECORD SHEETS G...
Page 114: ...Blank...
Page 117: ...OPTIONAL I O MODULES FOR AFU H...
Page 118: ...Blank...
Page 124: ...LONG PATHLENGTH OPTION App 1...
Page 126: ...Opacity and Dust Monitor User Guide Model 4500 MkIII App 1 2...