Remote control commands
R&S
®
NGU
141
User Manual 1179.2531.02 ─ 03
Example:
SENS:CURR:RANG? (@1)
Queries current range for measurement at channel 1.
SENSe:VOLTage:RANGe:AUTO
<arg0>[, <Channel list>]
SENSe:VOLTage:RANGe:AUTO?
[<Channel list>]
Sets or queries auto range for voltage measurement accuracy.
Parameters:
<arg0>
1
Enables auto range for voltage.
0
Disables auto range for voltage.
Parameters for setting and query:
<Channel list>
<list>
Example:
SENS:VOLT:RANG:AUTO 1
Enables auto range for voltage.
Example:
SENS:VOLT:RANG:AUTO? (@1)
Queries auto range state for voltage measurement accuracy at
channel 1.
SENSe:VOLTage:RANGe[:UPPer]
<arg0>[, <Channel list>]
SENSe:VOLTage:RANGe[:UPPer]?
[<Channel list>]
Sets or queries the voltage range for measurement. There is a selection of 20 V and 6
V range.
Parameters:
<arg0>
Defines the voltage range for measurement.
Default unit: V
Parameters for setting and query:
<Channel list>
<list>
Example:
SENS:VOLT:RANG 20
Sets the instrument to the 20 V measurement accuracy.
Refers to datasheet for the measurement accuracy in the 20 V
range.
Example:
SENS:VOLT:RANG? (@1)
Queries voltage range for measurement at channel 1.
Configuration commands
www.allice.de
Allice Messtechnik GmbH