Remote control commands
R&S
®
NGU
182
User Manual 1179.2531.02 ─ 03
.................................................................................................184
FLOG[:STATe]
<arg0>
FLOG[:STATe]?
Sets or queries the FastLog state.
Parameters:
<arg0>
1
Enables the FastLog state.
0
Disables the FastLog state.
Example:
FLOG 1
Activates the Ch1 FastLog state.
FLOG:DATA?
Queries FastLog data as a block.
The block is returned in the binary format starting in the sequence of voltage followed
by current measurements, i.e. V, I, V, I, ....
The R&S
NGU accepts the line message EOI and/or the ASCII character NL (0Ah) as
an indication that data transmission has been completed
The binary data stream must be concluded with EOI or NL or EOI followed by NL. If the
data stream is not concluded with either EOI or NL, the R&S
NGU will wait for addi-
tional data. In the case of a binary data transmission, the R&S
NGU ignores the bit
combination NL (0Ah) within the data stream.
The binary data block has the following structure:
#<LengthofLength><Length><block_data>
Example: #234<block_data>
●
<LengthofLength> specifies how many positions the subsequent length specifica-
tion occupies ("2" in the example)
Data logging commands