Remote control commands
R&S
®
NGU
185
User Manual 1179.2531.02 ─ 03
Example:
LOG ON
LOG?
-> 1
Data logging function is activated.
LOG:COUNt
<count>
LOG:COUNt?
[<Return min or max>]
Sets, queries the number of measurement values to be captured.
Setting parameters:
<count>
<numeric value> | MIN | MAX
<numeric value>
Number of measurement values to be captured is set in the
range of 1 to 10000000.
MIN
Minimum number of measurement values to be captured is set
at 1.
MAX
Maximum number of measurement values to be captured is set
at 10000000.
Parameters for setting and query:
<count>
MIN | MINimum | MAX | MAXimum
Returns the number of measurement values.
Example:
LOG:COUN MAX
LOG:COUN? MAX
-> 10000000
LOG:DURation
<span>
LOG:DURation?
[<Return min or max>]
Sets or queries the duration of the data logging.
Setting parameters:
<span>
<numeric value> | MIN | MAX
<numeric value>
Duration of the data logging captured in the range of 0 s to
3.49*10^5 s.
MIN
Minimum duration of the data logging captured at 0 s.
MAX
Maximum duration of the data logging captured at 3.49*10^5 s.
Default unit: s
Parameters for setting and query:
<span>
MIN | MINimum | MAX | MAXimum
Returns the duration of the data logging.
Example:
LOG:DUR MAX
LOG:DUR? MAX
-> 349000
Data logging commands