Remote control commands
R&S
®
NGU
170
User Manual 1179.2531.02 ─ 03
Parameters:
<arg0>
CC | CV | CR | SINK | PROTection | OUTPut
If "OUTPut" is selected, the "fault output" will be active if the out-
put of the selected channel is off.
Example:
DIO:FAUL:SOUR PROT
Sets the "operation modes" of the digital output fault source as
protection. If any of the protection modes (OCP, OCP, PPP, OTP
and Sense) of the digital output fault source is active, a trigger
event is triggered to the trigger-out signals on the instrument.
DIO:FAULt[:STATe]
<arg0>
DIO:FAULt[:STATe]?
Sets or queries the digital output fault.
Parameters:
<arg0>
1
Enables digital output fault.
0
Disables digital output fault.
Example:
DIO:FAUL 1
Enables the digital output fault.
DIO:FAULt:SIGNal
<arg0>
Select digital output fault signal type.
Parameters:
<arg0>
CONStant | PULSe
CONStant
A constant level trigger signal is sent out
PULSe
An output pulse of 100 ms trigger signal is sent out
DIO:OUTPut:SOURce
<arg0>, <arg1>
DIO:OUTPut:SOURce?
<arg0>
Sets or queries the digital output source.
Parameters:
<arg1>
OUTPut | TRIGger | FORCed
OUTPut
Selected channel output is used as the digital output source.
Advanced operating commands
www.allice.de
Allice Messtechnik GmbH