Remote control commands
R&S
®
NGU
191
User Manual 1179.2531.02 ─ 03
Parameters:
<Enable_Value>
Bit mask in decimal representation
Range:
0 to 65535
Example:
STATus:QUEStionable:INSTrument:ISUMmary1:
ENABle?
Queries the event register for the Standard QUEStionable Reg-
ister group.
STATus:QUEStionable:INSTrument[:EVENt]?
STATus:QUEStionable:INSTrument:ISUMmary<Channel>[:EVENt]?
Returns the contents of the EVENt part of the status register to check whether an
event has occurred since the last reading. Reading an EVENt register deletes its con-
tents.
Suffix:
<Channel>
.
1..n
Return values:
<Event>
Event bits in decimal representation
Range:
0 to 65535
Usage:
Query only
STATus:QUEStionable:INSTrument:NTRansition
<arg0>
STATus:QUEStionable:INSTrument:NTRansition?
STATus:QUEStionable:INSTrument:ISUMmary<Channel>:NTRansition
<NegativeTransition>
STATus:QUEStionable:INSTrument:ISUMmary<Channel>:NTRansition?
Sets or queries the negative transition filter. Setting a bit in the negative transition filter
shall cause a 1 to 0 transition in the corresponding bit of the associated condition regis-
ter to cause a 1 to be written in the associated bit of the corresponding event register.
Suffix:
<Channel>
.
1..n
Parameters:
<NegativeTransition> Range:
1 to 65535
Example:
STATus:QUEStionable:INSTrument:ISUMmary1:
NTRansition?
Query for negative transition.
Status reporting commands
www.allice.de
Allice Messtechnik GmbH