Remote control commands
R&S
®
NGU
157
User Manual 1179.2531.02 ─ 03
Example:
MEAS:ENER:RES (@1)
Resets the energy counter at channel 1.
Usage:
Setting only
MEASure[:SCALar]:ENERgy:STATe
<state>
MEASure[:SCALar]:ENERgy:STATe?
[<Channel list>]
Sets or queries the energy counter state for the selected channel.
Parameters:
<state>
1
Activates the energy counter.
0
Deactivates the energy counter.
Parameters for setting and query:
<Channel list>
<list>
Example:
MEAS:ENER:STAT ON
MEAS:ENER:STAT?
MEAS:ENER:STAT?
-> 1
Energy counter of Ch1 is enabled.
Example:
MEAS:ENER:STAT? (@1)
Queries the energy counter state at channel 1.
MEASure[:SCALar]:ENERgy:UNIT
<arg0>
MEASure[:SCALar]:ENERgy:UNIT?
Sets or queries the measured unit for energy.
Setting parameters:
<arg0>
WS | WH
WS
Energy in Watt per second.
WH
Energy in Watt per hour.
Example:
MEAS:ENER:UNIT WH
MEASure[:SCALar]:STATistic:COUNt?
[<Channel list>]
Queries the number of measurements.
Parameters:
<Channel list>
<list>
Example:
MEAS:STAT:COUN? (@1)
Queries the number of measurements at channel 1.
Usage:
Query only
Measurement commands